Why do I need to take two measurements at the same occasion with the Beacon Osstell?
Osstell ISQ Beacon
Status:
Open Jun 10, 2024 - 01:02 PM
1answer
Answer
Jun 10, 2024 - 01:12 PM
The SmartPeg measures the stability in all directions and singles out the most stable and the least stable direction. Sometimes, there is no difference between them. To make sure to measure both values, the probe should be held perpendicular to the jaw line for one measurement, and in line with the jaw line for the other measurement. The two ISQ’s represent these values. If the instrument finds them both at the same time, and they differ more than 3 ISQ units, they are presented on the display together.
View More Product Details
Osstell ISQ Beacon
The Osstell Beacon is an innovative and highly intuitive handheld device designed to assess dental implant stability using the Osstell ISQ and RFA technology. Determine Implant Stability in Seconds! Optimal ease of use and intuitive, color-coded ISQ measurements Cordless for Optimal transportability...